Recommended separation distances between portable and mobile RF communications equipment and the
The Flocare
®
Infinity
enteral feeding pump is intended for use in an electromagnetic environment in which radiated RF dis-
turbances are controlled. The customer or the user of the Flocare
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
Flocare
Infinity
as recommended below, according to the maximum output power of the communications equipment.
®
Rated Maximum output power
of transmitter
(W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in Watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
guidance and manufacturer's declaration – Electromagnetic immunity
The Flocare
Infinity
enteral feeding pump is intended for use in the electromagnetic environment specified below. The
®
customer or the user of the Flocare
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1
At 80 Mhz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measurement field strength in the location in which the Flocare
cable RF compliance above, the Flocare
mance is observed, additional measures may be necessary, such as reorienting or relocating the Flocare
b. Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
ZMNL900071_AW.indd 31
Flocare
Infinity
®
Separation distance according to frequency of transmitter (m)
150 KHz to 80 MHz
CD
d = 0,12
P
0,01
0,04
0,12
0,37
1,17
Infinity
enteral feeding pump should assure that it is used in such an environment.
®
IEC 60601 test level
Compliance level
3 Vrms
3 Vrms
150 KHz to 80 MHz
150 KHz to 80 MHz
3 V/m
30 V/m
80 Mhz to 2,5 GHz
80 Mhz to 2,5 GHz
Infinity
should be observed to verify normal operation. If abnormal perfor-
®
enteral feeding pump
Infinity
can help prevent electromagnetic interference
®
80 MHz to 800 MHz
CD
d = 0,12
P
0,01
0,04
0,12
0,37
1,17
Electromagnetic environment – guidance
Portable and mobile RF communications equi-
pment should be used no closer to any part of
the Flocare
®
Infinity
recommended separation distance calculated
from the equation applicable toe the frequency
of the transmitter:
Recommended separation distance
d = 1,17CD P
d = 0,12CD P
d = 0,23CD P
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in
each frequency range.
Interference may occur
in the vicinity of equipment
marked with the following
symbol:
Infinity
®
-29-
800 MHz to 2,5 GHz
CD
d = 0,23
P
0,02
0,07
0,23
0,74
2,33
, including cables, than the
80 MHz to 800 MHz
800 MHz to 2,5 GHz
a
b
is used exceeds the appli-
®
Infinity
.
11/7/11 4:35 PM