TECHNICAL SpECIFICATIONS
• This Flocare
®
Infinity
to EN 60601-1-2, EN 61000-3-2, EN 61000-3-3, RTCA DO-160D standards for electromagnetic emissions
and immunity and are in compliance with Directive 93/42/EEC.
0344
• IEC 601: Class II
• Microprocessor controlled
• BF Equipment
• Charger:
input 100-240V AC / 50-60 Hz / 0.4A Max.
output 5 V DC 2.4 A
• Battery: internal rechargeable Lithium ion battery, 3.6 V DC 2000 mAh
• Battery capacity: 24 hours at 125 ml/h
• Jet waterproof: IPX 5
• IPX 5: Splash water proof: Water jets from any direction shall have no effect on the pump.
• Dimensions: 140 x 95 x 35 mm
• Weight: approx. 392 g
• Accuracy flow rate: ±5,0% with appropriate Flocare
• Humidity:
• Temperature:
• Atmospheric pressure:
• Occlusion detection pressure: Upstream occlusion:
• Air bubble detection: The amount of air, that must pass the air sensor before the air alarm is activated, varies
from 0.5 to 1,5 ml which relates to an air bubble with an approximate length in the tubing of 9 - 26 cm.
• The Flocare
®
Infinity
• The use of other accessories, adaptors and cables than listed within this manual may result in increased
emissions or decreased immunity of the equipment of the Flocare
• Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided in the accompanying documents.
• Portable and mobile RF communications equipment (cellular telephones) can affect medical electrical equip-
ment. If the Flocare
be observed to verify normal operation.
• In case of pump scrapping, always notify your Nutricia sales unit of the pumps serial number. Pump scrap-
ping should always occur according to local legislation.
ZMNL900071_AW.indd 28
UK
™
pump in combination with the adapter and the Flocare
Medical Device Directive: Class IIa
Operation mode:
Storage:
Operation mode:
Storage and transportation: -20°C to +65°C
Operation mode:
Storage and transportation: 50-106 kPa
Downstream occlusion:
™
can safely be operated on commercial aircraft.
®
Infinity
™
pump is used adjacent to or stacked with other equipment, the pump should
®
Infinity
®
Infinity
™
pump set
30% to 75% noncondensing
10% to 95 % noncondensing
+5°C to +40°C
70-106 kPa
-34 kPa (tolerance 21 kPa)
83 kPa (tolerane 21 kPa)
Infinity
pump.
®
™
-26-
™