• Press the "START/STOP" key to pause the pump.
• Change the program by pressing the required key (DEBIT or DOSE=VOL) and adjust using the "+" or "-" keys.
• Restart the pump by pressing the "START/STOP" key again.
All parameters and values can be cleared one after the other in the following way:
• Press "START/STOP" to pause the pump.
• Select the parameter that needs to be cleared.
• Press the "EFFACER" key to clear the memory. The parameter or value returns to its default value:
ml/h = 0 ml/h
VOL = cont
INFO = 0 ml
In order to keep a clear overview of the daily amount of feed delivered, clear the total volume delivered as
each daily feeding schedule is started, as follows:
START/STOp
EFFACER
• When another feeding program needs to be started, or the pump will be used for another patient, all set-
tings (rate and dose) can be returned to default and the volume delivered can be set at zero by simply
pressing the "EFFACER" key and keeping it pressed for 2 seconds.
• The Flocare
• The battery symbol in the display is automatically replaced by the plug symbol when the adapter is
plugged in. The "fuel gauge" then indicates that the battery is charging by displaying the segments in
an ascending low to high pattern starting with the leftmost segment.
This pattern continuously repeats while the pump is charging.
TAL VOLUME DEL
TERVAL BOLUS
• In case of power failure the pump automatically switches to battery powered operation.
• When during functioning the last block is gone, the battery symbol will blink to indicate that there is approxima-
tely 1 hour of charge left. The display will flash "BATT" every 3 seconds, alternating with the active display and
the pump will beep every 2 seconds to remind the user of the low battery charge condition. Plug in the charger
to continue to run and recharge the battery.
• As the battery level decreases the pump may not be able to deliver tube feeds at rates greater than 300 ml/h,
even though some battery capacity remains. The pump will alarm and indicate "BATT".
If this happens, either recharge the battery or select a lower rate to complete the feed cycle.
ZMNL900071_AW.indd 23
TO CHANgE THE FEEdINg pROgRAM dURINg FUNCTIONINg
CLEARINg THE MEMORY
= No dose set. Pump will feed continuously until feed container is empty or pump
is stopped
= Volume delivered is cleared
• Switch the pump "ON"
• Immediately after the self-test the pump displays total volume delivered.
• Press the "EFFACER" key. The total volume delivered returns to "0 ml".
• The pump displays the previous programmed flow rate (ml/h).
• The pump is in the "HOLD" mode ready for programming.
Infinity
™
pump is equipped with a Lithium-ion battery.
®
• To check the status of the battery, disconnect the charger from the pump and turn the
pump on. The bars above the battery represents the "fuel gauge" of the battery. Each bar
is approximately 1/4 of a full battery charge. If 2 bars appear the battery is half full and
remaining operating time is approximately 12 hours at a flow rate of 125 ml/h.
BATTERY OpERATION