• Connect the set to the feed container as instructed on the packaging of the Flocare
Remove the dust cap from the step connector.
• The Flocare
Infinity
®
the set has no roller clamp.
Fill the pump set completely with feed by gently pinching on the side of the cassette that is marked with a
drop (point A on figure 5a) or fill the pump set with help of the pump (see paragraph: filling the pump set).
• Open the pump door by pressing up on the lower wall and rotating the door upwards at the same time
(see figure 5b).
• Position the looped section of the silicone tubing around the rotor.
Stretching lightly, (see figure 5c) seat the cassette into the pump (see figure 5d).
• Close the pump door.
START/STOp
The "hold" mode is used to temporarily stop the flow of feed:
• to change the feeding program (installed parameters),
• to change the feed container,
ZMNL900071_AW.indd 20
UK

OpERATINg INSTRUCTIONS

INSERTION OF THE FLOCARE
pump sets are equipped with an "automatic free-flow protection", as a consequence
SWITCHINg "ON"
For proper pump operation, make sure the pump door is closed prior to switching the
pump on by holding the
The pump beeps and carries out a short self test, showing the pump serial number
in 8 digits. Verify that all display segments and symbols are active as shown in figure
6 (see next page).
The pump displays the total volume delivered since the memory was cleared and
switches in the hold mode ready for programming.
SWITCHINg "OFF"
Press the
key and keep it pressed for 2 seconds. A continuous alarm will be heard
and the pump switches off.
The feeding program (installed parameters) and total volume administered since the
last clearance will be retained in the pump's memory.
If the pump shuts down due to low battery voltage, the memory will be retained for
24 hours.
"HOLd" MOdE
To temporarily pause the pump, or switch into "hold" mode whilst operating, press the
"START/STOP" key once. Three beeps are heard and the run symbol disappears.
The programmed flow rate (ml/h), volume (DOSE=VOL) and the administered volume
(ml) are retained.
INFINITY
pUMp SET
®
key for 2 seconds otherwise the pump will generate an error.
-18-
Infinity
pump set.
®
11/7/11 4:35 PM
A
P
S
T
a
T
T
T