Problem Solver Chart - Nutricia Flocare Infinity PMS2655U Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Using methods other than those described may cause the pump to function incorrectly.
Condition
NO SET
• The set is not fitted or wrongly fitted
• The pressure sensor area is dirty
pUSH STRT
• The pump has been untouched in
ENd OF dOSE
• The programmed dose/volume has
pROg
• No flow rate is installed.
dOOR
• The door is not correctly closed.
AIR
• The air sensor detected an empty
• The air sensor area is dirty.
• The feeding set is not inserted cor-
BATT
• The battery capacity is too low.
the pump
remains working
Battery, "E" and "F"
• Battery failure.
flashing
FILL SET
• The pump is priming the set.
OCC IN
• The pump detected an upstream
• The pressure sensor area is dirty
OCC OUT
• The pump detected a
• The pressure sensor area is dirty
ZMNL900071_AW.indd 25

pROBLEM SOLvER CHART

Always follow the instructions below in case a problem occurs.
Cause
in the pump.
(Fig 1, position 5 & 6).
hold mode for 3 minutes or more.
been administered.
Flow rate = 0 ml/h
pump set.
rectly (Fig 1, position 8).
The pump is not able to deliver
highly viscous fluids at high flow
rates with the present charge level
of the battery.
occlusion between the pump and
the feeding bag.
(Fig 1, position 5 & 6).
downstream occlusion between the
pump and the patient.
(Fig 1, position 5 & 6).
Correction
• Stop the alarm by pressing "START/STOP"
• Insert the Flocare
Infinity
®
indicated on the blister packaging (or refer to Fig 5a - 5d)
and close the door.
• Restart the pump.
• Clean the sensors, reinsert the feeding set in the pump
and restart the pump.
• Stop the alarm and prolong the hold mode with another 3
minutes by pressing "START/STOP" key.
• Program the pump and start it by pressing the "START/
STOP".
• Turn the pump off by pressing the
during 2 seconds or
• Clear the memory of the total volume administered (see
paragraph: clearing the memory), reprogram a new fee-
ding schedule and start the pump by pressing "START/
STOP".
• Make sure the correct flow rate is programmed.
• Make sure the door is properly closed prior to
starting a feeding program.
• Replace the empty feeding reservoir and continue fee-
ding. If necessary prime the set.
• Make sure the air sensor is clean.
• Make sure the feeding set is properly inserted in the
pump. (Fig 5a - 5d).
• Connect the adaptor to the pump and mains and charge
the pump for approx. 6 hours. During charging the pump
can be used.
• Turn the pump "off", contact your healthcare provider or
refer the pump to service.
• Press the "FILL SET" key another time to stop the pump
and bring it back in the hold mode.
• Stop the alarm by pressing the "START/STOP" key.
• Remove the feeding set out of the pump and check the
permeability by flushing the line.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the feeding set in the pump
and restart the pump.
• Stop the alarm by pressing the "START/STOP" key.
• Check the permeability of the feeding tube by
aspirating liquid via the medication port.
• Re-insert the feeding set in the pump and restart.
• Clean the sensors, reinsert the feeding set in the pump
and restart the pump.
-23-