• to administer medication without switching the pump off,
• to silence an alarm and correct problems.
After 3 minutes a two tone audible alarm sounds and the message "PAUSE" appears in the display.
Press "START/STOP" to stop the alarm and to extend the hold mode by a further 3 minutes or press "START/
STOP" twice (3 beeps will be heard) to resume programmed settings.
pURgER
When the pump is in the hold mode: Press and hold the "PURGER" key for 2 seconds to activate the
"PURGER" maneuver. Release the button as soon as the pump generates a beep and starts pumping at
a flow rate of approx. 700ml/h. During this maneuver the message "PURGER" appears in the display. The
pump will automatically stop when the Flocare
this "PURGER" maneuver can at anytime be stopped by pushing the "PURGER" key a second time. The
r
pump will return to the hold mode when the "PURGER" maneuver is complete or stopped.
d
The pump has a liquid crystal display (L.C.D.) with large alphanumeric characters, smaller symbols, words
and a back light.
The following information can be found in the display:
• Flow rate (ml/h), volume (ml), is displayed through the large characters.
d
Words below describe what the number relates to (rate, dose or volume).
The pump also displays messages, for example "end of dose" will appear when the pump has finished
delivering a single feed dose.
TOTAL VOLUME DEL
e
INTERVAL BOLUS
Figure 6
The back light of the display turns off 10 seconds after the last key is pressed.
The back light will switch on for 10 seconds when the charger is connected to the mains.
ZMNL900071_AW.indd 21
FILLINg THE pUMp SET
The "PURGER" function offered by the Infinity pump is used to fill a (new) feeding
set with nutrition (or water as the case may be). When the "PURGER" function runs
all alarms are deactivated helping avoid alarms being inadvertently triggered (e.g.
air alarm).
The "pURgER" function should only be used to fill an empty (air-filled) feeding set. Use of the
"pURgER" function at any other time may cause the pump to incorrectly calibrate. When the
"pURgER" function is initiated the pump will calibrate to the giving set.
Pack Infinity
®
dISpLAY
• The arcs rotate when the pump is running.
• The remaining battery capacity (see paragraph: battery
• The wall plug symbol indicates that the charger is plugged in.
pump set is completely filled with feed. However
™
operation).
-19-
UK