U93
Parameters
Impedance of each
bridge element
Response time
Excitation level
(nominal)
Excitation level
(minimum)
Excitation level
(maximum)
Calibration date
Calibration initials
Calibration period
(days)
Measurement location
ID
1)
Typical values for an HBM U93/2 kN type force transducer, manufactured on 1.6. 2006
The user can also write other templates, such as the Signal Conditioning
template.
A2081-2.2 en/de/fr
1)
Value
Unit
345.0
ohm
1.0000000
sec
u
5.0
V
0.5
V
12.0
V
1‐Jun‐2006
HBM
730
days
0
Rights
level
ID
Input resistance according to the
HBM data sheet
ID
Not relevant to HBM transducers.
ID
Nominal (rated) excitation voltage
according to the HBM data sheet
ID
Lower limit for the operating range of
the excitation voltage according to
the HBM data sheet.
ID
Upper limit for the operating range of
the excitation voltage according to
the HBM data sheet.
CAL
Date the manufacturing certificate
was created at HBM (or date of last
calibration or entry of TEDS data).
Format: day‐month‐year.
CAL
Initials of the calibrator or calibration
laboratory concerned (max. three
characters).
CAL
Time before recalibration, calculated
from the date specified under
Calibration Date.
USR
Identification number for the
measuring point (assigned by the
user). Possible values: 0 to 2047
(numbers only). The HBM Channel
Comment template is also available
as a supplement.
Explanation
HBM
19