1. Clean and anesthetize the skull pin sites.
2. Allow the local anesthetic to take effect.
Note
Integra recommends the use of an
antibacterial ointment on the tips of the
screws to help avoid infection.
3. Using the head ring wrench, insert and tighten two
diagonally-opposed head ring screws.
Caution: When installing the head ring screws:
• Only use Integra disposable head ring screws.
• Select the appropriate head ring screw length so that the screws protrude at
least 13 mm from the outside surface of the head ring posts.
• Use all four (4) head ring screws to attach the head ring.
Chapter 3 . Assembling the Universal Compact Head Ring . 18