Guidance and manufacturer's declaration - electromagnetic immunity for all EQUIPMENT
and SYSTEMS
Guidance and manufacturer's declaration - electromagnetic immunity
This camera is intended for use in the electromagnetic environment specified below. The
customer or the user of this camera should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Electrostatic
discharge (ESD)
±6 kV contact
IEC 61000-4-2
±8 kV air
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV line(s) to
line(s)
IEC 61000-4-5
±2 kV line(s) to
earth
Voltage dips,
<5 % U
T
short
(>95 % dip in U
interruptions and
for 0,5 cycle
voltage variations
on power supply
40 % U
T
input lines
(60 % dip in U
for 5 cycles
IEC 61000-4-11
70 % U
T
(30 % dip in U
for 25 cycles
<5 % U
T
(>95 % dip in U
for 5 sec
Power frequency
3A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Compliance level
Electromagnetic
environment – guidance
Floors should be wood,
±2, 4 & 6 kV
concrete or ceramic tile. If
contact
floors are covered with
synthetic material, the
±2, 4 & 8 kV air
relative humidity should
be at least 30 %.
±2 kV for power
Mains power quality
supply lines
should be that of a typical
commercial or hospital
±1 kV for input/
environment.
output lines
±0.5 & 1 kV
Mains power quality
differential mode
should be that of a typical
commercial or hospital
±0.5, 1 & 2 kV
environment.
common mode
<5 % U
Mains power quality
T
)
(>95 % dip in U
)
should be that of a typical
T
T
for 0,5 cycle
commercial or hospital
environment. If the user of
40 % U
this camera requires
T
)
(60 % dip in U
)
continued operation
T
T
for 5 cycles
during power mains
interruptions, it is
70 % U
recommended that this
T
)
(30 % dip in U
)
camera be powered from
T
T
for 25 cycles
an uninterruptible power
supply or a battery.
<5 % U
T
)
(>95 % dip in U
)
T
T
for 5 sec
3A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.