SPECIFICATION 
Accuracy is specified for a period of one year after calibration and at 23±5`C with relative humidity up to 75%. 
Accuracy specifications Lake the form of: 
± (1 % of Reading]+[number of Least Significant Digits]) 
 
Grounding Voltage 
RANGE  RESOLUTION 
200V 
 
0.1 V 
Max. allowable voltage: 200V 
Frequency range: 40Hz‐ 400Hz 
 
Grounding Resistance 
RANGE  RESOLUTION 
20 
 
0.01   
200   
0.1 
2000   
1.0 
 
Test current: about AC 2mA 
  
 
 
OPERATION INSTRUCTION 
 
1. Checking the Batteries 
 
lf the display shows'~" or the display is blank alter you switch the tester on, replace the batteries     
immediately. 
2. Testing the Grounding Voltage 
Waming: Don't test any voltage which is above AC200V. 
 
a. 
Set the function/range switch to "EARTH V" position. 
 
b. 
Connect the red voltage test lead (or the red normal‐test Iead ) to "V" terminal.  
 
                Connect the black voltage test Iead ( or the green normal‐test lead ) to "E" terminal. 
 
c. 
Connect the voltage test leads ( or the clips) to the two points between which   t he voltage is  
to be measured. 
 
d. 
Read the reading on the display. 
 
3. Testing the Grounding Resistance 
Waming: Thera is a voltage up to 50V between "E" terminal and "C" terminal, or between "E" terminal and "P" 
terminal. Don't touch the metal of the fest leads with finger or skin for safety. 
 
 
a. 
Before measuring, make sure the plugs of normal‐test leads have been well  
 
inserted into the specified terminals. lf the plug connections are loose, the measurement will be not     
accurate. 
 
b. 
Make sure the grounding voltage of the earth electrode 
  
 
to be measured is lass than 10V, otherwise the grounding resistance    
 
measurement will be not accurate. You can turn off the instruments which use the earth electrode to 
decrease the grounding voltage before measuring grounding resistance. 
 
c. 
The stabile of reading may be affected ff you do measurement near a large  
 
voltage transformer. You can turn off this voltage transformer beforehand or select another 
measurement position. 
 
 
 
 
 
 
 
 
ACCURACY 
 
± (1.5%+5) 
ACCURACY 
±(2.0%+3), or±0. 10 
 
±(2.0%+3) 
 
±(2.0%+3) 