34~42.2°C (93,2~108°F)
4. Accuracy level: ±0.2°C (0.4°F) be-
tween 35~42°C (95~107.6°F) and
±0.3°C (0.5°F) beyond this range
5. Operating range: 10~40°C (50 °F
~104 °F); relative humidity 15/85 %
6. The "Repeatability" is 0.17°C.
7. Battery life: around 3,000 readings.
Enable normal use: 1 year.
8. Expected Service Life: 4 years.
9. Enclosure Rating: IP22
10. Dimensions: 120.4 x 37.8 x 40.3
mm; Weight: 57.7 grams includ-
ing battery.
A. Storage
temperature
-20~+50°C, RH≤85%
B. Transportation temperature shall
be less than 70°C, RH≤95%
C. Atmospheric pressure: 800~1013 hPa.
The device should not submerge
into any liquids and expose it to direct
moisture.
INTENDED USE
The ear thermometer is electronic
thermometer using an infrared
detector (thermopile detector) to
detect body temperature from ear
canal in people of all ages.
INTENDED OPERATOR
Have eight years of education, no
maximum.
Ex.SN:E512A000001, the first "E" is
External, the second number "5"
is the last number of manufac-
ture year, the third and the fourth
number "12" is the manufacture
month, the others is the serial
number.
Classified
60601-1International Standards :
1. Internally powered appliance.
2.
Applied part: BF type.
3. Continuous operation mode
CE 1639: conforms to the
93/42/CEE + 2007/47/EC Medical
Equipment Standard.
Note:
1. The thermometer was produced
2. Portable and mobile radio commu-
range:
to
EN
21
in compliance with all standards
currently in force and meets all
magnetic compatibility protec-
tion requirements. However, it is
necessary to take certain EMC pre-
cautions and prepare and use the
thermometer following the EMC
instructions provided in this docu-
ment.
nication devices can affect the cor-
rect functioning of electro-medical
devices. In the presence of electro-
magnetic disturbance, it is highly.