8
Transmitter Prosonic Flow 93***-***********C
Terminals
20
+
Designation
Relay
Functional
60 V DC/100 mA
values
or
30 V AC/500 mA
Intrinsically
no
safe circuit
U
=
260 V AC
m
I
=
m
Transmitter Prosonic Flow 93***-***********D
Terminals
20
+
Designation
Status input
Functional
3...30 V DC
= 5 k Ω
values
R
i
Intrinsically
no
safe circuit
U
=
260 V AC
m
I
=
m
Transmitter Prosonic Flow 93***-***********L
Terminals
20
+
Designation
Status input
Functional
3...30 V DC
= 5 k Ω
values
R
i
Intrinsically
no
safe circuit
U
=
260 V AC
m
I
=
m
21
22
23
–
+
–
Relay
60 V DC/100 mA
or
30 V AC/500 mA
no
260 V AC
500 mA
21
22
23
–
+
–
Relay
60 V DC/100 mA
or
30 V AC/500 mA
no
260 V AC
500 mA
21
22
23
–
+
–
Relay
60 V DC/100 mA
or
30 V AC/500 mA
no
260 V AC
500 mA
PROline Prosonic Flow 93
24
25
26
+
–
+
Pulse/frequency
Current output
output
f = 2...10000 Hz
active: 0/4...20 mA
active: 24 V DC/25 mA
(max. 250 mA/20 ms),
R
L
>100 Ω
R
passive: max. 30 V DC
L
passive: 30 V DC/
250 mA
no
260 V AC
24
25
26
+
–
+
Pulse/frequency
Current output
output
f = 2...10000 Hz
active: 0/4...20 mA
active: 24 V DC/25 mA
(max. 250 mA/20 ms),
R
L
>100 Ω
R
passive: max. 30 V DC
L
passive: 30 V DC/
250 mA
no
260 V AC
24
25
26
+
–
+
Relay
Current output
60 V DC/100 mA
active: 0/4...20 mA
or
30 V AC/500 mA
R
L
passive: max. 30 V DC
no
260 V AC
Endress+Hauser
27
–
HART
< 700 Ω
R
L
HART ≥ 250 Ω
no
260 V AC
27
–
HART
< 700 Ω
R
L
HART ≥ 250 Ω
no
260 V AC
27
–
HART
< 700 Ω
R
L
HART ≥ 250 Ω
no
260 V AC