10. CALIBRATION
Turn it on again, and during the count from 9 to 0 press the key
The display will show "CAL" for a few seconds. While it shows "CAL", press the keys
in the calibration section. The display will show "unload".
Remove any weight that is placed on the platform. Press the key
The display will show the last calibration weight used. If this weight is correct you can keep pressing the key
the arrow keys to change the value of the calibration weight. When it will be correct press the key
The display will show "load". Place the calibration weight on the platform. Press the key
If the calibration is correct the display will return to normal mode, if an error message appears, try to recalibrate, because perhaps some
factor could impede proper calibration.
If the problem persists please contact Baxtran or its supplier. After calibration, the viewer should be reviewed to verify that the calibration and
linearity are correct. If necessary repeat calibration, especially make sure that the viewer is in a stable position before accepting any weight.

11. ERROR CODES

ERROR CODES
- - - - -
Err 1
Err 2
Err 4
Err 6
(Only personal authorized technician JUMPER K1)
Charging
C
DESCRIPTION
Above the range
Error when configuring the
date
Error when configuring time
Configuration zero error
Charging
Charging
A/D converter out of range
Charging
C
.
Charging
Charging
C
.
Charging
Charging
C
Charging
C
DETERMINATION
Remove the weight of the platform. If the problem persists contact
BAXTRAN or with your distributor for assistance.
Introduce the date using the correct format and the adequate values.
Formato: aa.mm.dd.
Introduce the date using the correct format and the adequate values.
Formato: hh.mm.ss.
The indicator was not correctly configured in the 0 range or it was
C
turned on while the
from the platform and press again.
C
Us the key
for set the display in 0 value.
If the problem persists please contact Baxtran or its supplier.
The values of the A/D converter are outside the normal range. Remove
the weight of the platform if there is an excess of weight. Indicates that
the load cell or electronics may not be functioning properly.
USER'S MANUAL BR40
Charging
C
C
,
and
In this same sequence to enter
C
.
.
key was pressured. Remove the weright
EN
EN
. If it is not correct, use
45