MEMORY OPERATIONS

This unit is equipped with advanced power off memory capability. It memorizes the last timer setting
once the cleaning cycle is started, and the last temperature setting once the heater is started.
1. Last timer setting recall.
a.
5 seconds after the timer counts down to 00:00, the last timer setting will
reappear. Press the On/Off key twice, it will repeat the last cycle with the same timer
setting.
b. Once the timer counts down to 00:00, touch the On/Off key within 5 seconds will
recall the last timer setting. Touch it again will repeat the last cycle with the same
timer setting.
c. If power is turned off during the cleaning cycle, once it is powered on again, the last
timer setting will reappear.
d. If the timer setting is changed during the cleaning cycle and the timer counts down
to 00:00, the new timer setting will reappear as in 1A, 1B or 1C.
e. If the timer setting is changed but the cleaning cycle is not started, it will stay until the
cleaning cycle is started. It will not stay if the power is turned off.
f.
If the On/Off key is touched twice during the cleaning cycle, the cycle will stop and
the timer setting will stop running. It'll resume the original cycle once the On/Off key
is touched twice again.
2. Last temperature setting recall.
a. The last temperature setting will remain once the heater is started.
b. If the temperature setting is changed but the heater is not started, the previous
temperature setting will remain after the unit is powered off and powered on again.
NOTE: To prevent accidental lockups of the touch-sensing control panel, please observe the
following:
1. Do no touch two or more keys at the same time.
Do not leave water or solution on the control panel.
2.
Do not push the control panel with too much force. Only light touch is needed.
3.
If the control panel is locked up, do the following to reset it. If the first step does not work, move on
to the next step.
1. Turn off the power switch.
Turn off the power switch multiple times.
2.
Unplug the power cord and wait for 5-10 minutes.
3.
Leave the unit unplugged for several hours.
4.
11