C C l l e e a a n n i i n n g g
W W A A R R N N I I N N G G
• Always unplug the product before you begin cleaning or disinfecting.
• Do not immerse the product in cleaning or disinfectant solutions.
• Do not spray or allow liquids to enter the power cord ports or the hose ports.
• Always make sure that you wipe each product with clean water and dry each product after you clean. Some cleaning
agents are corrosive in nature and may cause damage to the product. Failure to follow the cleaning and disinfecting
instructions may void your warranty.
C C A A U U T T I I O O N N
• Do not over expose the product to higher concentration chemical solutions as these may degrade the product.
• Failure to follow manufacturing instructions may also affect useful life of the product.
Recommended cleaning agents include:
A clean damp cloth and mild soapy water
To clean the external side of the pump:
1. Unplug the controller.
2. Wipe the outside of the product with a clean, soft cloth moistened with a mild soap and water solution to remove foreign
material.
3. Follow specified contact time in accordance with the manufacturer's instructions for use.
4. Using a clean, soft cloth, wipe the outside of the product with water.
5. Dry the external surface with a clean, dry cloth.
2874-009-005 Rev AA.0