If abnormal conditions appear on the screen during the test process, withdraw and reinsert the finger to
restore normal operation.
The device has normal useful life of three years from its first use.
The strap attached to the product is made from non-allergenic material. Stop using the strap in the event
that it causes reactions in the patient. Furthermore, pay attention to how the strap is used. Do not wear it
around the neck to avoid injuring the patient.
The instrument does not have a low battery alarm function; it only shows the low battery level. Change the
battery when its charge is depleted.
None of the parameters are monitored by an alarm function. Do not use the device in situations where
alarms are required.
Batteries must be removed before storing the device for more than one month to prevent leakage.
The two parts of the device are connected by a flexible circuit. Do not twist or pull on the connection.
2. Overview
The pulse oxygen saturation is the percentage of HbO
tration in the blood. It is an important respiratory bio-parameter. In order to improve the ease and accuracy of
SpO
measurements, our company has developed the Pulse Oximeter. The device is capable of simultaneously
2
measuring the pulse rate.
The Pulse Oximeter is a compact, energy-saving device that is both easy to use and portable. It only requires
the insertion of one finger into the fingertip photoelectric sensor for diagnosis. A display screen directly shows
the SpO
value measured.
2
34
Oxystart® - EN
in the total Hb in the blood, known as the O
2
concen-
2