13.6.1 Zero Calibration

13.6.2 Turbidity Calibration

36
Spectrophotometer NANOCOLOR
Clicking
opens the window for the zero calibration
é
ê
Select the desired cuvette size and start the calibration procedure
by confirming with
ment first and then asks you to insert the calibration cuvette.
Make sure that the calibration cuvette used is free of scratches
and turbidity. If necessary, carry out a turbidity measurement
before calibration (target value < 2 NTU). Clean the outside of
the cuvette with the enclosed cuvette cleaning cloth.
Insert the supplied calibration cuvette into the cuvette shaft; the
calibration is continued automatically.
The calibration can be interrupted by clicking the
When the calibration is finished, the date of the calibration is stored
in the instrument and displayed the next time the calibration pro-
cedure is started.
Clicking the icon
the instrument. Select the desired cuvette size. Start the calibra-
tion procedure by confirming with
asks for the corresponding standards and starts the respective
measurement automatically. When the calibration is complete, the
date of the calibration is stored in the instrument and displayed
the next time the calibration procedure is called up.