4 Battery Charging
The Orion3 contains Li-ion type rechargeable batteries. Please read and comply
with the safety information. Failure to do so may cause the battery to become
hot, explode, or ignite and cause serious injury.
Please advise the user of the charging information in this section.
The batteries are not field replaceable and will be replaced as necessary when
serviced.
If the Orion3 is subjected to temperatures lower than -10°C ( 14°F) it must be
returned to be inspected for possible damage to the battery pack.
Charging
Socket
1. The battery should only be charged in temperatures between 10˚C to 40˚C
(50˚F to 104˚F).
2. Only charge the batteries in the Orion3 using the charger supplied with
the product.
Do not use this charger for charging any other device.
3. The battery charger can be used with an electrical outlet that is rated
between 100V and 240V AC, 50/60Hz.
4. Never plug the charger into the Orion3 while still wearing the knee.
5. During charging no power is delivered to the Orion3 control system.
6. The knee module may become warm during charging but should never be
too hot to touch. There should be no unusual smell, if there is immediately
switch OFF the battery, unplug the charger from the electrical supply and
contact your service provider.
7. Do not refit the knee until the charger has been disconnected.
8. Always recharge the Orion3 before use if it has not been used for an
extended period of time.
5
Battery
On/Off
Switch
[+], [-]
Buttons
and LED
938375/1-0816