Ametek mocon Dansensor MAP Mix ProVectus Guide Utilisateur page 46

Table des Matières

Publicité

44
Guide utilisateur
Location Register write Encoding
0x0050
0x0051
0x0066
P/N 300864-H
01/2020
Set mixer type
1 byte unsigned. Saved in non volatile memory.
(Write only)
0 = Not used in MMP (Ignored)
1 = CO
2 = N
2
3 = CO
4 = N
2
NOTE! It is NOT allowed to send 2-gas modes to a 3-gas
mixer or vice versa
Set input range
1 byte unsigned. Saved in non volatile memory
for voltage input.
0 = 0 to 10V
0-10/2-10V
1 = 2 to 10V (default)
NOTE! If V-in (voltage input) is not used, always select the
range 2 to 10 V and short circuit the input.
Output valve
1 byte unsigned.
mode (write)
See also read 0xAD06
0 = Stop mixer
1 = Start mixer
FR
Dansensor
/N
2
2
/O
2
/O
2
2
/CO
/O
2
2
MAP Mix Provectus
®