Dansensor
MAP Mix Provectus
®
Location Register read
0xA07E
Mix 1 actual
setting
(Read only)
0xA07F
Mix 2 actual
setting
(Read only)
0xA080
Mix 3 actual
setting
(Read only)
0xA084
Actual Vacuum
buffer pressure
0xA06E
Vacuum buffer
set pressure
(read only)
0xA072
Vacuum buffer
hysteresis
(read only)
0xA076
Vacuum buffer
min on time
(times 100ms)
(read only)
0xA077
Vacuum buffer
min off time
(times 100ms)
(read only)
0xA0D6
Output valve
mode
FR
Encoding
1 unsigned byte
Range 0 to 100%
For a 3-gas mixer this is the mixer setting of %CO
For a CO
/N
mixer this is the mixer setting of %N
2
2
For a N
/O
mixer this is the mixer setting of %N
2
2
For a CO
/O
mixer this is the mixer setting of %CO
2
2
1 unsigned byte
Range 0 to 100%
For a 3-gas mixer this is the mixer setting of %N
For a CO
/N
mixer this is the mixer setting of %CO
2
2
For a N
/O
mixer this is the mixer setting of %O
2
2
For a CO
/O
mixer this is the mixer setting of %O
2
2
1 unsigned byte
Range 0 to 100%
Only present in a 3-gas mixer.
This is the mixer setting of %O
For a 2-gas mixer this register returns zero.
4 byte float
Actual pressure in the vacuum buffer [bar]
4 byte float
Actual set point for the pressure in the vacuum buffer.
Typical 3.0 or 5.0 bars
4 byte float
Typical 0.3 bar. When buffer pressure reach "Vacuum buffer
set pressure" - "Vacuum buffer hysteresis" the valve is
turned on. When the pressure reaches "Vacuum buffer set
pressure" it is turned off.
1 byte unsigned
Not used, will always return 0
1 byte unsigned
Not used, will always return 0
1 byte unsigned
See also write 0x0066
0 = OFF, mixer is stopped
1 = Running (Auto). Normal operation mode.
Guide utilisateur
2
2
2
2
2
2
2
2
.
2
P/N 300864-H
01/2020
41