Dansensor
MAP Mix Provectus
®
Read only commands (R "Read memory (Operational commands")
Location Register read
0xA065
Status register
(Read only)
0xA067
Error register
(Read only)
FR
Encoding
2 bytes unsigned integer
Each bit represent a status condition
0x0001:
Always 0 as MMP is already initialized.
0x0002:
A system FAULT has occurred
LED on mixer is blinking red at 2Hz (fast
blink)
Possible error causes are:
•
Flow cannot be maintained (to low
input capacity flow/pressure or
backpressure to high)
•
Any ERROR listed in User Guide
After FAULT situation, the mixer must be
restarted by either changing the mixture,
sending stop/start sequence using
command 0x0066 or external signal
0x0004:
Not used in MMP
0x0008:
One or more gas input pressure is below
limit.
LED on mixer is blinking red at 2Hz (slow
blink)
0x0010:
Gas inlet pressure is OK.
For mixer type CO
Gas 2 (N
For all other mixer types this bit is set
when Gas 1inlet pressure is OK.
0x0020:
Gas inlet pressure is OK.
For mixer type CO
Gas 1 (CO
For all other mixer types this bit is set
when Gas 2 inlet pressure is OK.
0x0040:
Gas 3 inlet pressure is OK. If the mixer is
configured as a 2-gas mixer this bit is 1
(pressure OK)
0x0080:
This bit is 1 when mixer is running and
GAS is flowing through the mixer.
0x0100:
Not used in MMP
0x0200:
Not used in MMP
0x0400:
Not used in MMP
2 bytes unsigned integer
Not used in MMP, this register is always 0x0000 in MMP.