Guidance and manufacturer's declaration-electromagnetic immunity
The pM-T01K is intended for use in the electromagnetic environment specified below.
The customer or the user of the pM-T01K should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
± 6 kV contact
(ESD) IEC 61000-4-2
± 8 kV air
Electrical fast tran-
± 2kV for power sup-
sient/burst IEC 61000-
ply lines
4-4
± 1kV for input/out-
put lines
Surge IEC 61000-4-5 ± 1kV line(s) to line(s)
± 2kV line(s) to earth
Voltage Dips, short
<5% UT(>95% dip in
interruptions and
UT) for 0,5 cycle
voltage variations on
40% UT(60% dip in
power supply input
UT) for 5 cycles
lines IEC 61000-4-11
70% UT(30% dip in
UT) for 25 cycles
<5% UT(>95% dip in
UT) for 5 s
Power frequency
3 A/m
(50/60 Hz) magnetic
field IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Recommended separation distance between portable and mobile RF communications equipment and the pM-T01K
The pM-T01K is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the pM-T01K can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the pM-
T01K as recommended below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter / W
150 kHz to 80 MHz /
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
All manuals and user guides at all-guides.com
Compliance
level
± 6 kV contact
± 8 kV air
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
Not applicable
3 A/m
Separation distance according to frequency of transmitter / m
80 MHz to 800 MHz
d=1,2√P
/ d=1,2√P
N/A
0,12
N/A
0,38
N/A
1,2
N/A
3,8
N/A
Electromagnetic environment-guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidi-
ty should be at least 30%
Mains power quality should be that of a typical com-
mercial or hospital environment.
Mains power quality should be that of a typical com-
mercial or hospital environment.
Mains power quality should be that of a typical com-
mercial or hospital environment. If the user of the pM-
T01K series requires continued operation during power
mains interruptions, it is recommended that the pM-
T01K series be powered from an uninterruptible power
supply or a battery.
The pM-T01K power frequency magnetic fields should
be at levels characteristic of a typical location in a typical
commercial or hospital environment.
800 MHz to 2,5 GHz / d=2,3√P
12
0,23
0,73
2,3
7,3
23
39