Table 2
For all me equipment and me systems
Guidance and manufacturer's declaration – electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity
BM55 is intended for use in the electromagnetic environment specified below. The customer or the user of BM55 should assure that it is
used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
Power frequency (50/ 60 Hz)
magnetic field IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5 % U
(> 95 % dip in U
)
T
T
for 0,5 cycle
40 % U
(60 % dip in U
)
T
T
for 5 cycles
70 % U
(30 % dip in U
)
T
T
for 25 cycles
< 5 % U
(> 95 % dip in U
)
T
T
for 5 sec
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV input/output lines
± 1 kV differential mode
± 2 kV common mode
< 5 % U
(> 95 % dip in U
)
T
T
for 0,5 cycle
40 % U
(60 % dip in U
)
T
T
for 5 cycles
70 % U
(30 % dip in U
)
T
T
for 25 cycles
< 5 % U
(> 95 % dip in U
)
T
T
for 5 sec
3 A/m
86
Electromagnetic environment – guidance
Floors should be wood, concrete or cera-
mic tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Mains power quality should be that of a typi-
cal commercial or hospital environment.
Mains power quality should be that of a typi-
cal commercial or hospital environment. If
the user of BM55 requires continued ope-
ration during power mains interruptions, it
is recommended that BM55 is be powered
from an uninterruptible power supply or a
battery.
Power frequency magnetic fields should be
at levels characteristic of a typical location
in a typical commercial or hospital envi-
ronment.