TECHNICAL SPECIFICATIONS
• This Flocare
Infinity
®
are designed to EN 60601-1-2, EN 61000-3-2, EN 61000-3-3, RTCA DO-160D standards for
electromagnetic emissions and immunity and are in compliance with Directive 93/42/EEC.
0344
• IEC 601: Class II
• Microprocessor controlled
• Type BF Shock Protection (EN60601-1) degree of protection against electrical shock. No electrical
connection to the patient. Drop from any height of 90cm shall not damage pump operation.
• Charger:
input 100-240V AC / 50-60 Hz / 0.4A Max.
output 5 V DC 3 A
• Battery: internal rechargeable Lithium ion battery, 3.7 V DC 2500 mAh
• Battery capacity: ∼24 hours at 125 ml/h for a new battery.
• IPX 5: EN60529 degree of protection against water entering the enclosure. Water jets from any
direction shall have no effect on the pump.
• Dimensions: ∼140 x 95 x 35 mm
• Weight: approx. 392 g
• Accuracy flow rate: ±5,0% with appropriate Flocare
• Humidity:
• Temperature:
• Atmospheric pressure:
• Occlusion detection pressure: Upstream occlusion:
• Air bubble detection: The amount of air, that must pass the air sensor before the air alarm is activa-
ted, varies from 0.5 to 1,5 ml which relates to an air bubble with an approximate length in the tubing
of 9 - 26 cm.
• The Flocare
Infinity
®
• The use of other accessories, chargers and cables than listed within this manual may result in incre-
ased emissions or decreased immunity of the equipment of the Flocare
• Medical electrical equipment needs special precautions regarding EMC and needs to be installed
and put into service according to the EMC information provided in the accompanying documents.
• Portable and mobile RF communications equipment (cellular telephones) can affect medical electri-
cal equipment. If the Flocare
the pump should be observed to verify normal operation.
• In case of pump scrapping, always notify your Nutricia sales unit of the pumps serial number.
• At the end of their service lifetime, the pump and its electrical accessories (AC Adapter Charger
and Data Download Accessory Cable), should be disposed of according to local standards and
regulations governing the disposal of electronic waste (e-waste). The pump contains a rechar-
geable lithium-ion battery. When disposing of the pump, be sure to discard this equipment in a
manner consistent with local policy for expired battery operated equipment. Other acces-
sories can be disposed of or recycled as standard non-hazardous waste.
ZMNL900908_AW.indd 31
pump in combination with the charger and the Flocare
™
Medical Device Directive: Class IIa
Operation mode:
Storage:
Operation mode:
Storage and transportation: -20°C to +65°C
Operation mode:
Storage and transportation: 50-106 kPa
Downstream occlusion:
can safely be operated on commercial aircraft.
™
Infinity
pump is used adjacent to or stacked with other equipment,
®
™
Infinity
pump set
®
™
30% to 75% noncondensing
10% to 95 % noncondensing
+5°C to +40°C
70-106 kPa
-34 kPa (tolerance 21 kPa)
83 kPa (tolerane 21 kPa)
-29-
UK
Infinity
Nurse Call
®
™
Infinity
pump.
®
™