all settings (rate and dose) can be returned to default and the volume delivered can be set at
zero by simply pressing the "EFFACER" key and keeping it pressed for 2 seconds.
• The Flocare
mate 2 to 5 years. Performance may degrade due to age, excessive temperatures and number of
charging cycles.
• The battery symbol in the display is automatically replaced by the plug symbol when the charger is
plugged in. The "fuel gauge" then indicates that the battery is charging by displaying the bars in an
ascending low to high pattern starting with the leftmost bar.
TAL VOLUME DEL
TERVAL BOLUS
ning operating time is approximately 12 hours at a flow rate of 125 ml/h for a new battery.
• In case of power failure, when the pump is connected to mains, the pump automatically switches to
battery powered operation.
• When during functioning the last bar is gone, the battery symbol will blink to indicate that there is
approximately 1 hour of charge left. The display will flash "BATT" every 3 seconds, alternating with the active
display and the pump will beep every 2 seconds to remind the user of the low battery charge condition. Plug
in the charger to continue to run and recharge the battery.
• In case of battery failure, caused by, for example, an excessive temperature condition (which may resolve
itself with time) or by a complete battery failure, the battery symbol will flash. Turn the pump "off" and
contact your healthcare provider or refer the pump to service.
The power connector at the side of the pump offers the option of connecting the pump to an external
alarm system or Patient Data Monitoring System (PDMS).
• The Flocare
system. This external alarm system may be with open or closed contact. Follow the instructions for use
of the Flocare
• The Flocare
system. Follow the instructions for use of the Flocare
• For good pump operation, the pump should be kept clean and free of debris like (dried) tube nutrition. It
is recommended to clean the pump with a wet cloth after every finished feeding regimen and specifically
in between patient uses (if applicable).
• Detergent and disinfectant cleaning agents and wipes can damage plastic surfaces of medical devices
when they are not compatible with the surface material. Damaged surfaces may compromise the ability
to decontaminate medical devices adequately and / or may interfere with device function and affect their
integrity and performance. Ensure cleaning solutions, detergents and disinfectant wipes are compatible
with the device. Look for signs of damage, cracking, colour changes, etc. to the medical device during
and after the application. In case of doubt, contact your local (Nutricia) sales representative.
• Avoid harsh or agressive cleaners/ disinfectants such as acetone, benzene or iodine.
• Do not sterilize the pump by irradiation, EtO, steam or in an autoclave. Always unplug the pump prior
to cleaning to avoid electric shock hazard.
• On a regular basis thoroughly clean all surfaces of the pump (including the sensors and rotor) with warm
soapy water, a 5% bleach solution in water, a multipurpose disinfectant cleaner or a soft cloth.
ZMNL900908_AW.indd 26
UK
Infinity
pump is equipped with a Lithium-ion battery. The expected battery life is approxi-
®
This pattern continuously repeats while the pump is charging.
• To check the status of the battery, disconnect the charger from the pump and turn the
pump on. The bars above the battery represents the "fuel gauge" of the battery. Each bar is
approximately 1/4 of a full battery charge. If 2 bars appear the battery is half full and remai-
CONNECTION TO AN EXTERNAL ALARM OR PDMS SYSTEM
Infinity™ Nurse Call (corporate code 35752) enables connection to an external alarm
®
Infinity™ Nurse Call to connect the pump.
®
Infinity™ PDMS Cable (corporate code 35776) enables connection to an external PDMS
®

MAINTENANCE AND CLEANING

BATTERY OPERATION

®
Infinity™ PDMS Cable to connect the pump.