SET UP MODE

The set up mode is used
This set up mode may only be entered by healthcare professionals and authorised personnel trained to use this application of
r
the pump.
The patient or his/her relatives without permission of the physician, dietician, nurse or other licensed practitioner, may not
change the settings of the set up mode.
2 sec
Hold the '+' key down and press
First the alarm level can be set:
"BIP FORT" (+) or "BIP BAS" (-)
Use the "+" or "-" key to change the setting from "fort" to "bas" or from "bas" to "fort".
ZMNL900908_AW.indd 29
Cause
• Calibration not yet completed
on the current feeding set
• Only the current feeding
schedule is allowed for this
patient. Another feeding pro-
gram is not allowed by your
healthcare professional.
• The self test detected an electro-
nic error
• Keypad pressed or stuck during
pump start up.
• The wall outlet doesn't work.
• The charger is damaged.
• to set the alarm level,
• to lock the keypad,
• to switch the audible alarm off when a dose is administered.
• to switch the light of the LCD permanently on when connected to
the mains.
• to enable connection with a Nurse call or Patient Data
Monitoring System (PDMS).
+
+
Bip fort (Beep high)
&
d
-
Bip bas (Beep low)
2 sec
Correction
• Stop the alarm by pressing the "START/STOP" key.
• Start the pump by pressing the "START/STOP" key
and let it run only briefly.
• Stop the pump by pressing the "START/STOP" key
insuring that there has been no occlusion out alarm.
• Remove the feeding set from the pump and reinsert
the feeding set in the pump.
• Restart the pump by pressing the "START/STOP"
key.
• The programming feature is blocked in the set up
mode of the pump. Ask your healthcare professional
to modify this setting.
• Turn the pump "off", make sure the pump door is
closed and switch the pump back "on". If the error
persists, contact your healthcare provider or refer
the pump to service.
• Restart pump by ensuring keypad is not pressed.
• Connect the pump to another wall outlet.
• Contact your healthcare professional or
Nutricia subsidiary to replace the charger.
TO SET THE ALARM LEVEL
for 2 seconds. The pump enters the set up mode.