PURGER
d
-
Hold the '+' key down and press 'PURGER' for 2 seconds. The pump enters the set up mode. Press
the "PURGER" key to switch to the next setting:
"NON VERROU" the keypad (-) or "VERROU" the keypad (+). In the "VERROU" mode the feeding
program of the pump can no longer be changed.
Use the "+" or "-" key to change the setting from "verrou" to "non verrou" or vice versa.
+
BIP FIN DOSE (Beep high)
PURGER
d
-
NO BIP (Beep low)
Hold the '+' key down and press 'start/stop' for 2 seconds. The pump enters the set up mode. Press
the "PURGER" key to switch to the next setting:
"BIP FIN DOSE" (+): the pump will give an alarm when the dose is administered or "NO BIP" (-):
the pump will not give an alarm after administering the dose. Use the "+" or "-" key to change the
setting from "BIP FIN DOSE" to "NO BIP" or vice versa.
+
PURGER
d
-
Hold the '+' key down and press 'start/stop' for 2 seconds. The pump enters the set up mode.
Press the "PURGER" key to switch to the next setting: "LUM ON" (+) the light of the LCD remains
on when the pump is connected to the mains. "LUM OFF" (-) the light of the LCD switches off after
a few seconds even if connected to the mains. Use the "+" or "-" key to change the setting from "LUM
ON" to "LUM OFF" or vice versa.
+
OUTP PDMS
PURGER
d
-
OUTP NRSE
Hold the '+' key down and press 'start/stop' for 2 seconds. The pump enters the set up mode.
Press the "PURGER" key to switch to the next setting: "OUTP PDMS" (+) to connect with a Patient
Data Monitoring System. "OUTP NRSE" to connect with a NURSE CALL system. Use the "+" or "-"
key to change to setting from "OUTP PDMS", to "OUTP NRSE" or "OUTP OFF" and vice versa.
Press the
key and hold it down for 2 seconds to exit the SET UP mode. The settings are auto-
matically saved.
ZMNL900908_AW.indd 30
UK
TO LOCK THE KEYPAD
Verrou (Locked)
Non verrou (Unlocked)
TO MUTE WHEN DOSE DONE
TO SWITCH LIGHT ON
LUM ON
d
LUM OFF
2 sec
TO CONNECT WITH NURSE CALL OR PDMS
d
2 sec
-28-