7. Cleaning
Remove general contamination with a soft cloth damped in oil free oxygen
compatible soap water and rinse with clean water. Used cleaning solution shall
comply with oxygen cleanliness standards for oxygen devices.
Disinfections can be carried out with an alcohol-based solution (spray or wipes).
If other cleaning solutions are used, check that they are not abrasive and that
they are compatible the product materials (including labels) and gas.
• Do not use cleaning solutions containing ammonia!
!
• Do not immerse in water or any liquid.
!
• Do not expose to high temperature (such as autoclave).
!
8. Maintenance
8.1 Product Life time
Serial number and date of production
Form of nine digit serial number stamped on the product is following:
YY MM XXXXX
YY: year of production
MM: month of production
XXXXX : sequence number
Example: serial number 090300521 shows the regulator produced in March
2009, with sequence number 521.
Product life time and waste management
Maximum life time of the product is 10 years.
At the end of the product's life time, the product must be withdrawn from
service (max.10 years). The owner shall put in place a relevant procedure to
ensure the product cannot be used again.
8.2 Repairs
Repairs
Repairs activities cover the replacement of the following damaged or missing
components:
• Quick connector and outlet connector
• Piston
The repairs shall be carried out by a person only authorized by the
manufacturer.
Any product sent back to the authorized person for maintenance shall be
properly packaged. The purpose of the maintenance has to be clearly specified
(repair). For product to be repaired a short description of fault and any reference
to a claim number might be helpful.