Guidance and Manufacturer's Declaration – Electromagnetic emissions
The Propulse is intended for use in the electromagnetic environment specified below. The customer
or user of the Propulse should assure that it is used in such an environment.
Emission tests
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class A
CISPR 11
Harmonic
Class A
emissions IEC
61000-3-2
Voltage
Complies
fluctuation/flicker
emissions IEC
61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Propulse is intended for use in the electromagnetic environment specified below. The customer
or the user of the Propulse should assure that it is used in such an environment
Immunity test
IEC 60601 test level
Electrostatic
6 kV contact
discharge (ESD) IEC
8 kV air
61000-4-2
Electrical fast
2 kV for power
transient/ burst
supply lines
IEC 61000-4-4
1 kV for
input/output lines
Surge
1 kV differential
IEC 61000-4-5
mode
2 kV common
mode
Voltage dips, short
< 5 % UT (> 95 %
interruptions and
dip in UT) for 0.5
voltage variations
cycle 40 % UT (60
on
% dip in UT) for 5
power supply input
cycles 70 % UT (30
lines
% dip in UT) for 25
IEC 61000-4-11
cycles
< 5 % UT (> 95 %
dip in UT) for 5 s
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the A.C. mains voltage prior to application of the test level.
Electromagnetic environment – guidance
The Propulse uses only energy for its internal function.
Therefore, its RF emissions are not likely to cause any
interference in nearby electronic equipment.
The Propulse is suitable for use in all establishments
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Compliance level
Electromagnetic environment -
guidance
6 kV contact
Floors should be wood, concrete
8 kV air
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
2 kV for power
Mains power quality should be
supply lines
that of a typical commercial or
Not applicable
hospital environment.
1 kV differential
Mains power quality should be
mode
that of a typical commercial or
2 kV common
hospital environment.
mode
< 5 % UT (> 95 %
Mains power quality should be
dip in UT) for 0.5
that of a typical commercial or
cycle 40 % UT (60
hospital environment. If the user
% dip in UT) for 5
of the Propulse requires
cycles 70 % UT
continued operation during power
(30 % dip in UT)
mains interruptions, it is
for 25 cycles
recommended that the Propulse
< 5 % UT (> 95 %
be powered from an
dip in UT) for 5 s
uninterruptible power supply or a
battery.
Not applicable
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.