Télécharger Imprimer la page

Medtronic Guardian Connect MMT-7821 Manuel D'utilisation page 28

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 35
Disposal
Discard the transmitter according to local regulations for battery disposal, or contact your
healthcare professional for disposal information.
Specifications
Biocompatibility
Applied parts
Operating conditions
Storage conditions
Battery life
Transmitter frequency
Maximum output pow-
er (EIRP)
Operating range
Transmitter expected
service life
Transmitter wireless communication
Quality of service
The transmitter and app connect via a BLE network. The transmitter sends glucose data
and system related alerts to the app. The transmitter and the app verify the integrity of
received data after wireless transmission. Quality of the connection is in accordance with
the Bluetooth Specification v4.0.
Transmitter: Complies with EN ISO 10993-1
Transmitter
Sensor
Transmitter temperature: 0 °C to 45 °C (32 °F to 113 °F)
Caution: When operating the transmitter on a tester in air temperatures
greater than 41 °C (106 °F), the temperature of the transmitter may ex-
ceed 43 °C (109 °F)
Transmitter relative humidity: 10% to 95% with no condensation
Transmitter pressure: 57.60 to 106.17 kPa (8.4 to 15.4 psi)
Charger temperature: 10 °C to 40 °C (50 °F to 104 °F)
Charger relative humidity: 30% to 75% with no condensation
Transmitter temperature: -20 °C to 55 °C (-4 °F to 131 °F)
Transmitter relative humidity: up to 95% with no condensation
Transmitter pressure: 57.60 to 106.17 kPa (8.4 to 15.4 psi)
Charger temperature: -10 °C to 50 °C (14 °F to 122 °F)
Charger relative humidity: 10% to 95% with no condensation
Transmitter: Six days of continuous glucose monitoring immediately fol-
lowing a full charge
Charger: The charger uses one new AAA battery to charge the transmitter
2.4 GHz band, Bluetooth Smart (version 4.0)
0.1 mW (-9.9 dBm)
Up to 6.1 meters (20 feet)
The transmitter expected service life is one year depending on patient us-
age.
-22-

Publicité

loading