Dry the transmitter and tester with a clean, dry cloth.
11
Prepare a 1:10 bleach solution by using one (1) part 8.25% bleach to nine (9) parts
water, for a final concentration of 0.8%. Make sure to prepare a fresh solution for
each use.
12
Make sure that you have completed the previous cleaning steps before disinfection.
With the tester still attached, soak the transmitter in the bleach solution for 20
minutes.
13
Rinse the transmitter under running room temperature tap water for three minutes.
14
Place the transmitter and tester on a clean, dry cloth and air dry them completely.
-16-