WARNING
The appliance needs special precaution regarding EMC and needs to be installed and put into service
f
f
according to the EMC information provided in the chapter "Normative reference".
The appliance needs to be installed and put into service in accordance with the information provided in the
f
f
accompanying documents.
Wireless communications equipment such as wireless home network devices, mobile phones, cordless
f
f
telephones and their base stations, walkie-talkies can affect this applicable and should be kept at least a
distance d away from the appliance. The distance d is calculated by the manufacturer from the 800 MHz to 2.5
GHz column of table in the chapter "Normative reference.
The Ultra Nomade is intended for use in the electromagnetic environment specified below. The customer or the user of
the Ultra Nomade should assure that it is used in such an environment.
IMMUNITY
IEC 60601 TEST LEVEL
TEST
Electrostatic
±6 kV contact
discharge (ESD)
IEC 61000-4-2
±2 kV for power supply lines
and dc lines
Electrical fast
transient/burst
IEC 61000-4-4
±1 kV for input/output lines
±1 kV line(s) and neutral
Surge
IEC 61000-4-5
± 2 kV line(s) to earth
<5 % U
for 0,5 cycle
Voltage
dips, short
40 % U
interruptions
and voltage
variations
70 % U
on power supply
for 25 cycles
input lines
IEC 61000-4-11
<5 % U
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
EMC INFORMATION
NORMATIVE REFERENCES
GUIDANCE AND MANUFACTURER'S DECLARATION
ELECTROMAGNETIC IMMUNITY
COMPLIANCE LEVEL
±8 kV air
±2 kV for power supply
Not applicable for input/
±1 kV line(s) and neutral
N/A(§7.2 EN61000-4-5)
(>95 % dip in U
)
<5 % U
T
T
(60 % dip in U
)
40 % U
T
T
for 5 cycles
(30 % dip in U
)
70 % U
T
T
(>95 % dip in U
)
<5 % U
T
T
for 5s
3 A/m
NOTE - U
is the a.c. mains voltage prior to application of the test level
T
- EMC INFORMATION
Floors should be wood, concrete or ceramic tile.
±6 kV contact
If floors are covered with synthetic material, the
±8 kV air
relative humidity should be at least 30 %.
lines
Mains power quality should be that of a typical
output lines
(>95 % dip in U
)
T
T
for 0,5 cycle
Mains power quality should be that of a typical
(60 % dip in U
)
T
T
commercial or hospital environment. If a dips or an
for 5 cycles
interruption of mains power occurs, the current of
the Ultra Nomade may be dropped off from normal
(30 % dip in U
)
T
T
level, it may be necessary to use uninterruptible
for 25 cycles
(>95 % dip in U
)
T
T
for 5s
Power frequency magnetic fields should be at
3A/m
levels characteristic of a typical location in a typical
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE
commercial or hospital environment.
power supply or a battery.
commercial or hospital environment.
51