Aesculap
Acculan 4 Charger GA320
6.
Working with the device
6.1
System set-up
Connecting to the power supply
Risk of death by electric shock!
►
Connect the product only to a grounded power
supply.
DANGER
►
Carry out a visual inspection before connecting the product to the
mains power:
– Check the mains power cord for any damage.
– Check the charger for any damage (e.g. bent, short-circuited con-
tacts of the charging bays).
►
Do not use the product if it is damaged, defective or dirty.
►
Set aside the product if it is damaged.
►
Clean product if dirty, see Validated reprocessing procedure.
►
Plug the mains power cord into the device connector 7.
►
Plug in the mains plug at the mains power socket.
After being connected to the mains power, the charger performs a self-
test. During this time the "Power On" 3 indicator and the
indicator fields for charging bays 2 must be illuminated on the front
panel of the device.
►
Carry out a visual inspection of the "Power On" indicator 3 and the
indicator fields for charging bays 1.
The indicator fields for the charging bays 2 go out and only
the "Power On" 3 indicator remains lit.
The charger is ready.
Note
In case of any deviation from the sequence described above, see Trouble-
shooting list
Putting out of operation
Note
The Acculan 4 Charger GA320 does not have a power ON/OFF switch. The
all-pole disconnection of the product from the main power supply is only
guaranteed when the power cord/mains connection line is unplugged.
►
Unplug the power cord from the mains power socket.
►
Unplug power cord from the device connector 7.
6
Acculan 4
®
6.2
Safe operation
Risk of injury and material damage due to inappro-
priate use of batteries!
►
Only insert batteries that are indicated in the
WARNING
Intended Use.
►
Insert batteries in their correct position.
Risk of injury and material damage due to damaged
batteries!
►
Check batteries for any damage before charg-
WARNING
ing.
►
Do not use or recharge damaged batteries.
Total battery failure can occur by leaving the bat-
teries in the charger while disconnected from the
power supply!
CAUTION
►
Always leave charger connected to the power
supply. Indicator lamp for readiness must illumi-
nate.
►
Always remove batteries from the charger when
it is disconnected from the power supply.
Loss of capacity/performance due to prolonged
storage of discharged batteries!
►
After a successfully completed charging process,
CAUTION
leave the battery in the operational charger
(trickle charging).
►
As excessive battery temperatures will prolong the charging time,
allow batteries to cool down prior to charging.
►
To remove the battery, if necessary use a cloth as the batteries become
heated during the charging process.