Acculan 4 Charger GA320
7.4
Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
Risk of electric shock and fire hazard!
Unplug the power cord before cleaning.
DANGER
Do not use flammable or explosive cleaning or
disinfecting solutions.
Ensure that no fluids will penetrate the product.
Damage to, or destruction of the product caused by
mechanical cleaning/disinfection!
Only clean and disinfect the product manually.
CAUTION
Do not sterilize the product under any circum-
stances.
Risk of damage to the product due to inappropriate
cleaning/disinfecting agents!
Only use cleaning/disinfecting agents approved
CAUTION
for surface cleaning. Follow the manufacturer's
instructions for the respective cleaning/disin-
fecting agent.
Do not introduce any fluid into the charging
bays.
7.5
Wipe disinfection for electrical devices without sterilization
Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe the specified application time (1 min minimum).
Clean the contacts in the charging bay with isopropanol or ethyl alco-
hol and a cotton swab. Do not use corrosive chemicals.
8
Acculan 4
®
T
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
Water quality
Chemicals
-
Meliseptol Wipes sensitive