Aesculap
Acculan 4 Charger GA320
7.4
Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
Risk of electric shock and fire hazard!
►
Unplug the power cord before cleaning.
DANGER
►
Do not use flammable or explosive cleaning or
disinfecting solutions.
►
Ensure that no fluids will penetrate the product.
Damage to, or destruction of the product caused by
mechanical cleaning/disinfection!
►
Only clean and disinfect the product manually.
CAUTION
►
Do not sterilize the product under any circum-
stances.
Risk of damage to the product due to inappropriate
cleaning/disinfecting agents!
►
Only use cleaning/disinfecting agents approved
CAUTION
for surface cleaning. Follow the manufacturer's
instructions for the respective cleaning/disin-
fecting agent.
►
Do not introduce any fluid into the charging
bays.
7.5
Wipe disinfection for electrical devices without sterilization
Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
►
Remove any visible residues with a disposable disinfectant wipe.
►
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
►
Observe the specified application time (1 min minimum).
►
Clean the contacts in the charging bay with isopropanol or ethyl alco-
hol and a cotton swab. Do not use corrosive chemicals.
8
Acculan 4
®
T
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
Water quality
Chemicals
-
Meliseptol Wipes sensitive