EN
4-4 Error Codes
Display
Error code
EC, E11 Stability error
The balance cannot stabilize due to an environmental
problem. Check around the pan. Prevent vibration, drafts,
temperature changes, static electricity and magnetic fields,
from influencing the balance. Refer to the separate
instruction manual "2-3 During Use".
To return to the weighing mode, press the
Out of the setting range
The data to be stored is out of the setting range.
Malfunction of the internal memory element of the
balance
If this error continues to be displayed, please request repair.
EC, E16 Internal mass error
Applying the internal mass does not yield a change in the
mass value as specified.
Confirm that there is nothing on the pan and perform the
weighing operation from the beginning again.
EC, E17 Internal mass error
The internal mass application mechanism does not function
properly. Perform the weighing operation from the beginning
again.
Abnormality in the internal memory data of the balance
If this error continues to be displayed, please request repair.
Abnormality in the internal memory data of the balance
If this error continues to be displayed, please request repair.
EC, E20 Sensitivity adjustment weight error (Positive value)
The sensitivity adjustment weight is too heavy. Confirm the
mass value of the weight. Press the CAL key to return to
the weighing mode.
EC, E21 Sensitivity adjustment weight error (Negative value)
The sensitivity adjustment weight is too light. Confirm the
mass value of the weight. Press the CAL key to return to
the weighing mode.
Overload error
A sample beyond the balance weighing capacity has been
placed on the pan. Remove the sample from the pan.
Weighing pan error
The mass value is too light.
Confirm that the weighing pan is properly installed and
adjust the sensitivity of the balance.
Description
11
CAL key.