RTD Type
Ni120 (672)
Cu10
Cu50
Cu100
YSI400
Pt100 (385)
Pt200 (385)
Pt500 (385)
Pt1000 (385)
Pt385-10
P1385-50
Pt100 (3926)
R100 (391 6)
Read Accuracy is based on 4-wire input. For 3-wire input add ± 0.05Ω assuming all three
RTD leads are matched.
Table 17: RTD Read and Source
Range (°C)
-80.0 ... 260.0
-100.0 ... 260.0
-180.0 ... 200.0
-180.0 ... 200.0
15.00 ... 50.00
-200.0 ... 100.0
100.0 ... 300.0
300.0 ... 600.0
600.0 ... 800.0
-200.0 ... 100.0
100.0 ... 300.0
300.0 ... 630.0
-200.0 ... 100.0
100.0 ... 300.0
300.0 ... 630.0
-200.0 ... 100.0
100.0 ... 300.0
300.0 ... 630.0
-200.0 ... 100.0
100.0 ... 300.0
300.0 ... 600.0
600.0 ... 800.0
-200.0 ... 100.0
100.0 ... 300.0
300.0 ... 600.0
600.0 ... 800.0
-200.0 ... 100.0
100.0 ... 300.0
300.0 ... 630.0
-200.0 ... 100.0
100.0 ... 300.0
300.0 ... 630.0
Accuracy (°C)
0.2
1.4
0.4
0.3
0.1
0.2
0.3
0.4
0.5
0.8
0.9
1.0
0.4
0.5
0.6
0.2
0.3
0.4
1.4
1.6
1.8
2.0
0.4
0.5
0.6
0.7
0.2
0.3
0.4
0.2
0.3
0.4