Les langues disponibles

Read (mV)
Source (mV)
Maximum current output in voltage ranges is 1mA with an output impedance of
TC Type
J
K
T
E
R
S
B
C
XK
BP
L
U
N
All TC errors include CJC errors
CJC errors outside of 23 ± 5 °C is 0.05°C/°C (In °C add .2 for cold junction compensation error.)
86
Table 15: Thermocouple Measurement / Source
Range
-10.000mV - 75.000mV
-10.000mV - 75.000mV
Table 16: Thermocouple Read and Source
-210.0 ... 0.0
800.0 ... 1200.0
-200.0 ... 0.0
0.0 ... 1000.0
1000.0 ... 1372.0
-250.0 ... 0.0
-250.0 ... -100.0
-100.0 ... 1000.0
0.0 ... 1767.0
0.0 ... 1767.0
600.0 ... 800.0
800.0 ... 1000.0
1000.0 ... 1820.0
0.0 ... 1000.0
1000.0 ... 2316.0
-200.0 ... 800.0
800.0 ... 2500.0
-200.0 ... 0.0
-200.0 ... 0.0
-200.0 ... 0.0
0.0 ... 1300.0
Accuracy(% of reading ± digit)
0.02% ± 10
0.02% ± 10
Range (°C)
0.0 ... 800.0
0.0 ... 400.0
0.0 ... 800.0
0.0 ... 900.0
0.0 ... 600.0
< -
Accuracy (°C)
0.4
0.2
0.3
0.6
0.3
0.5
0.6
0.2
0.6
0.2
1.2
1.2
1.2
1.3
1.5
0.6
2.3
0.2
0.9
2.3
0.25
0.2
0.5
0.25
0.8
0.4

Publicité