steAM steRILIZAtIon

Wrap as indicated and/or required.
Place in sterilizer which has been either pre-heated (by running empty through a short 10 minute run
at normal steam pressure) or has just been unloaded from a previous sterilizing load.
Sterilize at 15 lbs. (6.80 kg), 250°F (121°C), for 30 minutes ONLY. Watch time carefully. Do not leave in
for more than 30 minutes.
At the end of drying cycle, de-pressurize sterilizer immediately and open door minimally (just a small
crack is best).
Leave this instrument pack in partially opened sterilizer for at least 15 minutes. Do not remove until
instrument is cool to the touch. This is very important!
Place on a towel or open rack, but do not place upon any hard, cold surface that can trap and
condense moisture on the bottom surface of the pack.
DO NOT DRY-HEAT STERILIZE.
If available, gas autoclaving is preferable
NEVER IMMERSE INSTRUMENT, EVEN TO COOL. THIS WILL RESULT IN THE UNIT BEING DAMAGED
AND NEEDING REPAIR.

eto steRILIZAtIon

Precondition dermatome in chamber to at least 35% humidity at the selected sterilization temperature
for 30 minutes.
Dermatome can be sterilized wrapped or unwrapped.
Chamber temperature should not exceed 131°F (55°C) (99°F (37°C) is preferable).
Gas exposure time: as needed for validated sterilization (minimum 60 minutes at 131°F(55°C) or 180
minutes at 99°F (37°C)).
After exposure, evacuate chamber to at least 3 inches (7.62 cm) (230 m bar) vacuum to evacuate gas
from chamber.
Aerate chamber for at least 75 minutes with a minimum of 3 inches (7.62 cm) (900 m bar) of vacuum.
Chamber temperature must be maintained during aeration cycle.