Key
The HR-AZ series can perform
sensitivity adjustment using the built-
in weight.
In the function table mode, it works
as cancel key.
The stable weighing data is outputted
according to the function table.
In each operation procedure, it works
as determination key.
The current measurement value is regarded as zero and zero is displayed.
This function performs the sensitivity adjustment of the HR-AZ series using the built-in
weight (the internal mass).
Procedure
Step 1.
Connect the AC adapter to the balance and warm up the balance for at least one
hour with nothing on the weighing pan.
Step 2.
Place the accessory breeze break on the balance.
Press the CAL key to display
Step 3.
The sensitivity adjustment is performed using the internal mass.
Do not allow vibration or drafts to affect the balance.
Step 4.
If you choose the parameter to use the GLP output of the function table, a
sensitivity adjustment report is outputted after the sensitivity adjustment.
Step 5.
The balance will automatically return to weighing mode after sensitivity
adjustment.
This function performs the sensitivity adjustment using an external weight.
Procedure
Step 1.
Connect the AC adapter to the balance and warm up the it for
at least one hour with nothing on the weighing pan.
English (EN)
When pressed
When pressed and held (for 2 seconds)
The sensitivity adjustment mode is
performed.
No function at factory settings.
"Title block" and "End block" is
outputted according to GLP/GMP
settings on the function table.
.
5