Product Description - Linde Domistar Notice D'utilisation

Détendeurs haute pression
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

EN. Domistar, Hospistar, Neostar, PF4 - High Pressure regulators. Instruction for Use.
In case of storage at a temperature below -20 °C, do not operate the regulator
until it has been allowed to increase its temperature to a minimum of -20 °C.
At regulators intended for use with mixture of medical gases O
operation temperature limit +5°C. During normal use the flow outlet and
pressure outlet will sometimes have a frosty appearance. This is a normal
physical reaction in the valve, due to that the gas is going from high pressure
to low pressure (Joule Thompson effect). Ensure that all equipment that is
connected to the valve by at least a 2 metre hose.
4. Personnel instructions and training
According to Medical Devices Directive 93/42/EEC the product owner must ensure
that all personnel handling the product are provided with the operating
instructions and performance data and are fully trained to carry out that operation.
Trainees need to be supervised by an experienced person.

5. Product description

The regulator acts as a pressure-reducer, gas from the cylinder valve passes
through the pressure regulator to the user outlets.
A
Inlet stem
Regulator is fitted to the medical cylinder valve by mean of an inlet stem. The stem
can be bull nose type (male thread), nut type (female thread) or pin index type.
The inlet stem includes a filter.
B
Pressure gauge
A cylinder contents pressure gauge is provided to indicate the cylinder contents
(the medical cylinder valve shut-off valve must be set to the ON position to allow
the pressure to register on the gauge).
05
/N
O is lower
2
2