4025/4026/4261 CALIBRATION

This scale has been factory calibrated and does not require calibration prior to use.
maintenance, the scale can be calibrated. Calibration of your ProPlus
1.
Make sure the display is turned off, then press and hold the
HOLD/RELEASE button while pressing the ON/OFF or I/O
button.
Using the  keys and/or the numeric keypad, set the calibration
2.
load to at least half the capacity of the scale (for the best results
and greater accuracy, use the full capacity of the scale.
3.
Make sure there is nothing on the scale, then press the
ENTER button.
4.
Please wait 1-3 seconds, without touching or shaking the
scale, to allow the Zero Calibration to finish.
5.
No action required by operator.
6.
Load the scale with the required weight and press the ENTER
button. Please wait 1-3 seconds, without touching or shaking
the scale, to allow the Calibration to finish.
If calibration is successful, the display will show "Factor OK!".
7.
If the calibration process failed, it will show "Factor Fail!".
8.
If successful, the display will begin to reboot.
9.
The display will return to the start up message.
10. Please wait until the scale resumes normal operation and
then remove the weight from the scale.
Operator Action
14
®
scale is performed using pounds.
Display
"Enter load weight:
"400.0 lb"
"Enter load weight:
"400.0 lb"
Zero calibration:
Please clear the scale
Zero calibration:
Please wait
Weight calibration:
Put: XXX.X
Weight calibration
Please wait
Factor Update...
"Factor OK!"
or "Factor Fail!"
Rebooting...
Please wait
Health o meter
ProPlus
Health o meter
ProPlus