2.7.2 Emergency stop (Eagle 625)
2.7.3 Lowering in an emergency (Eagle 620)
2.7.4 Manual emergency lowering (Eagle 625)
WARNING: Risk of injury – Adjust the manual emergency lowering for the weight

of the patient.
1
Please note that the manual emergency lowering is only possible when the patient is sitting
in the patient hoist.
Handle pull force: 10 mm for force from 80 à 100N.
In emergencies, and during assembly and disassembly,
the emergency stop is activated by pressing the red
button on the control box.
Push on the button and turn in the direction of the arrows.
If the button is fully pressed to the back turn back in the
opposite direction of the arrows.
The emergency stop is deactivated by
turning the red button in the direction
of the arrows.
If the manual operation mechanism should be damaged, it
is possible to lower the cantilever arm from the control box.
The arm moves down for as long as the bottom button on
the control box panel is pressed.
Emergency lowering if the power fails or the batteries are
empty can be done using the red pull button c at the
lower end of the motor. This emergency lowering is
factory-set for a patient weight of 75 kg.
Standard calibration: 3000N, 16 mm/s to lower the
patiënt.