• To open the lithium battery compartment press at the textured
surface and simultaneously slide cover back
• Remove the lithium battery, grip lithium battery plug by the wire
and pull it gently out of the socket.
• Grip the plug of the new lithium battery by the wire and push it
irmly into the socket until you can hear it is snapping in.
• Insert new battery and close lithium battery compartment.
8
INSTRUCTION MANUAL - GEBRAU
MANUAL DE FUNCIONAMIENTO
Motorized Pipette Filler - Motorisierte Pipettierhilfe -
Pipeteur motorisé - Pipeta motorizada

7 - Maintenance

• Cleaning
1. Pull out the pipette.
2. Unscrew the adapter support.
3. Pull out the membrane ilter.
4. Remove the adapter from the adapter support by pushing it
upwards. The check value remain in the adapter.
5. To clean adapter with built in check valve, carefully rinse it
by using an appropriate compatible solvent (e.g. water) in a
wash bottle. Blow out the liquid and allow drying completely.
6. Place the new membrane ilter with the thicker end facing
downwards into the adapter (with the smaller conical side
up).
7. The entire pipette adapter including membrane ilter is
autoclavable at 121 ° C.
8. Reassemble the instrument in reverse order and carry out
leak test.
• Storage
The Fisherbrand Motorized Pipette shall be stored in a dry place.
The allowable storage temperate: -20°C to +50°C.
• Note:
Do not store the pipetting aid with a iller pipette
www.eu.fishersci.com/fisherbrand