Important Safety Instructions

Statement of Intended Use
Tripp Lite Medical-Grade UPS Systems are intended to support and protect non-medical
computer equipment and medical devices that require leakage current reduction, surge
protection, voltage regulation, line noise filtering and battery backup during power outages and
generator testing, both inside and outside patient care vicinities. Tripp Lite's Medical-Grade UPS
Systems come with hospital-grade plugs and receptacles that reduce leakage to below 100µA.
SAVE THESE INSTRUCTIONS
This manual contains important instructions that should be followed during the installation,
operation and storage of all Tripp Lite UPS Systems. Failure to heed these warnings will void
your warranty.
Note: UL 60601-1 certified UPS systems must not be used in countries other than the United States and
Canada and must be used only in healthcare facilities on grounded systems where conditions of
maintenance and supervision ensure that only qualified persons will service the electrical distribution system.
UPS Location Warnings
• Relevant symbols
Denotes that information in the manual should be reviewed before use.
Used to signal as a warning that a statement is particularly important and could pose
a safety risk.
Denotes that instructions should be followed for proper use.
Warns that users should not sit or step on the unit.
Indicates that you should not push the unit.
• Do not use this equipment within oxygen-enriched atmospheres, or within 1 ft. (0.3 m) of a
point at which an oxygen-enriched atmosphere is intentionally vented.
• Use caution when lifting UPS. Because of the considerable weight of all UPS systems, at
least two people should assist in lifting and installing them.
• Install your UPS indoors, away from excess moisture or heat, dust or direct sunlight.
• For best performance, the UPS should be used in a location that meets the following
conditions: Temperature: 32 to 104° F (0 to 40° C); Humidity: 0 to 95% (non-condensing);
Elevation: <6,652 ft. (2,000 m) above sea level.
• Leave adequate space around all sides of the UPS for proper ventilation. Do not obstruct its
vents or fan openings.
• Do not mount unit with its front or rear panel facing up or down (at any angle). Mounting in
this manner will seriously inhibit the unit's internal cooling, eventually causing product
damage not covered under warranty.
UPS Connection Warnings
• The UPS contains its own energy source (battery). The output terminals may be live even
when the UPS is not connected to an AC supply.
• Connect your UPS to a properly grounded AC power outlet. Do not modify the UPS's plug in
a way that would eliminate the UPS's connection to ground. Do not use adapters that
eliminate the UPS's connection to ground.
• Do not plug your UPS into itself; this will damage the UPS and void your warranty.
16-07-007-933607.indb 2
2
8/31/2016 10:04:04 AM