processor or waterproof wearing solution while underwater. If the
cable becomes accidently detached during water use, remove the
headpiece and sound processor from the water and dry both devices
prior to re-connecting the cable to the sound processor.
It is important to have the correct magnet strength, so you do not
experience discomfort or retention issues. If magnet strength is
insufficient, the headpiece may fall off more than is acceptable.
If magnet strength is excessive, you may experience irritation or
discomfort. Consult your cochlear implant professional if there are
any concerns regarding magnet strength. If deemed appropriate, an
audiologist may change the strength of the magnet in the headpiece.
Do not change magnet strength unless under the direction of a
cochlear implant professional. If you experience any redness,
irritation, or discomfort, discontinue use of the headpiece immediately
and contact a cochlear implant professional. See the headpiece
instructions for use for additional information regarding adjustment of
headpiece magnet strength.
Store additional headpieces away from items with magnetic strips
(e.g., credit cards, hotel room keys, etc.) as they may de-magnetize
cards.
The magnetic field created by the headpiece magnet may alter the
programming of magnetic shunts. Users of magnetic shunts should
consider off-the-shelf retention methods.
The digitally-coded, inductive transmission technology used in this
device is extremely reliable and experiences virtually no interference
from other devices. It should be noted, however, that when operating
the device near a computer terminal or other strong electromagnetic
field (e.g. RFID system, metal detectors, electromagnetic anti-theft
systems), it may be necessary to be at least 24" (60 cm) away to
ensure proper operation. If the Sky CI M90 sound processor does
not respond to the implant device because of an unusual field
disturbance, move away from the disturbing field.
The Sky CI M90 sound processors and accessories should be used
in accordance with the electromagnetic compatibility information
provided in the Guidance and Manufacturer's Declaration section of
your instructions for use document.
Although electromagnetic emissions of your Sky CI M90 sound
processor and accessories have been confirmed within safe limits,
some other devices could be sensitive to such emissions. If you
notice other devices do not behave as expected when your sound
processor is nearby, separate the devices.