DJO ARTROMOT-E2 Mode D'emploi page 80

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 94
Note!
• For an exclusive extension/flexion
protocol, motor B for the rotation
movement must be turned off as
described in steps 6 through 8 below.
• Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
6. Press the MENU key on the program-
ming unit to access level 3 (M3).
7. Press the Motor B ON/OFF parameter
key (
) to activate the parameter.
8. Deactivate motor B by pressing the
parameter key Motor B ON/OFF again
or the – (minus) key. The check mark
in the circle next to the symbol must
have disappeared.
9. Save the settings with the STOP key,
set the carriage to the starting posi-
tion with the START key and initiate
therapy by pressing the START key
again.
5.4.2 Isolated pronation/
supination
1. Complete the mechanical settings and
adjust the carriage to the patient as
described in section 4.2.
2. Press the MENU key on the program-
ming unit to access level 1 (M1).
3. Activate the parameter key
extension (
) or flexion (
then press the + / - keys to set the
carriage to the appropriate position
to be maintained during the isolated
pronation/supination movements.
4. Next set the desired range of motion
for pronation/supination by first press-
ing the pronation parameter key
(
), then set the values with the
+ / – keys.
5. Set the supination value (
same way.
Note!
• For an exclusive rotation protocol,
motor A for extension/flexion must
be turned off as described in steps 6
through 8 below.
• Having programmed the ranges of
motion, you can program additional
options, such as pause or speed.
6. Press the MENU key on the program-
ming unit to access level 3 (M3).
7. Press the Motor A ON/OFF parameter
key (
8. Deactivate motor A by pressing the
parameter key Motor A ON/OFF again
or the – (minus) key. The check mark
in the circle next to the symbol must
have disappeared.
9. Save the settings with the STOP key,
set the carriage to the starting position
with the START key and initiate therapy
by pressing the START key again.
),
) in the
75
) to activate the parameter.