Warning!
m
Patient hazard –
− Ensure an anatomically correct
setup of the CPM device suitable for
the patient to be treated. Therefore,
carefully verify the following settings/
positions:
(also refer to the numbers on the
device):
1. Horizontal adduction/abduction
(horizontal extension/flexion)
2. Height adjustment
3. Adjustment of the positioning angle
4. Adjustment of the forearm length
5. Axis adjustment motor A and
motor B
6. Backrest adjustment
7. ROM adjustment
− It is not permitted to change the
adjustment of 1 through 6 while the
CPM device is applied on a patient.
– Movements must not cause any pain
or irritation.
– Patients must be fully conscious
while being instructed in the use of
the CPM device and during therapy.
– The choice of the therapy param-
eters to program and of the therapy
protocols to use is restricted to the
responsible physician or therapist.
It is the physician's or therapist's deci-
sion whether or not to use the CPM
device on a specific patient.
− The patient must be familiar with the
functions of the ARTROMOT®-E2/-E2
compact programming unit and the
unit must be within easy reach of the
patient, allowing him or her to stop
therapy, if needed. Patients unable
to operate the programming unit,
e.g. paralytic patients, must never be
left unattended during therapy.
– Write the patient's name on the
patient chip card. The card should
only be used for this patient.
If the patient chip card is used for
another patient, be sure to delete the
previous patient's data from the
card first (see sections 4.1 and 5.3,
para graph "New Patient").
Use original chip cards only.
− All accessories used with the
ARTROMOT®-E2/-E2 compact must
be approved by DJO.
− Modifications to the medical device
described in this document without
the manufacturer's written consent is
prohibited.
− Ensure that the medical device is
securely set up on solid ground during
use.
− Do not allow parts of the body or any
objects (such as blankets, cushions or
cables) to get caught in the moving
parts of the CPM device.
− Always seat the patient on a chair
with four legs and without armrests
when using the ARTROMOT®-E2
compact.
Warning!
m
− Extreme caution should be taken when
in use around small children and babies!
Sufficient distance to the device is
mandatory for their safety!
− Never leave the device unattended
when it is switched on! Switch the
device off and disconnect the power
cord from the wall outlet!
− After use, store the device in a safe place!
Ensure device stability also during storage!
56