7. Possible Malfunctions
Problem
The digital display
doen not show
any symbols
Signallization of
low AC voltage.
Weighing
operation data
readout error.
Incorrect
calibration
Displayed error codes
Error
Meaning
code
• The calibration load inconsistent
with the display indication has been
used during the calibration.
no
•
The number of pieces inconsistent
with the display indication has
been used during the calibration.
•
Too low batteries voltage.
AC
• Calibration data loss.
•
Err-0
Damaged digital scale or remote
display.
• The load excesses the maximum
value.
• Calibration data loss.
Err-1
• Damaged digital scale or remote
display.
Cause
• Incorrectly connected
supply plug.
•
Discharged batteries.
•
Low batteries voltage.
• The "0" value has not
been set befor weighing.
•
The digital scale is not
calibrated.
•
The calibration load is not
precise.
Solution
•
Connect the supply plug
correctly.
•
Replace the batteries.
• Replace the batteries.
• Remove all objects from the
weighing surface, press
and begin the weighing
operation.
•
Calibrate the scale.
•
Use the precise calibration
load.
Solution
• Use the load corresponding to
the displayed value.
•
Use the number of pieces
corresponding to the displayed
value.
•
Replace the batteries or use the
mains adapter.
• Recalibrate.
•
Contact the service company.
• Remove the objects from the
weighing surface.
• Recalibrate.
• Contact the service company.
- 31 -