14
Resistance measurement (same for all models)
Maximum capacity load:
Maximum inductive load:
Conditions of guaranteed accuracy: After zero adjustment has been performed
Overload protection:
Overload current:
Resistance measurement (Measurement value/MAX/MIN/AVG)
(Accuracy guarantee
Range
range)
Ω
Ω
600.0
(0.0
to 600.0
Ω
Ω
6.000 k
(0.000 k
to 6.000 k
Ω
Ω
60.00 k
(0.00 k
to 60.00 k
Ω
Ω
600.0 k
(0.0 k
to 600.0 k
Auto range movement threshold: 6000 counts or more for upper range, 540 counts or less for lower range.
10 mF
10 H
Lower of 1700 V DC/1000 V AC or 2×10
30 mA or less at steady state, 1.5 A or less at transient state
Measurement
Resolution
current
Ω
Ω
)
0.1
200 µA±20%
Ω
Ω
) 0.001 k
100 µA±20%
Ω
Ω
)
0.01 k
10 µA±20%
Ω
Ω
)
0.1 k
1 µA±20%
Accuracy Table
7
V • Hz (energized for 1 minute)
Measurement accuracy
Ω
±0.7% rdg.±0.5
Ω
±0.7% rdg.±0.005 k
Ω
±0.7% rdg.±0.05 k
Ω
±0.7% rdg.±0.5 k
Open terminal
voltage
2.0 V DC or
less
2.0 V DC or
less
2.0 V DC or
less
2.0 V DC or
less
65