Operation And Maintenance - Lindab LT 600 Mode D'emploi

Masquer les pouces Voir aussi pour LT 600:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

12. Software
Lindab PC software is included as part of the package, and allows a PC to be used for data transmission and admin-
istration
You can use the software to create customers and measuring points/sequences in advance and upload them to the LT
600
The software can also be used to install updates of the LT 600 rmware as well as the PC software itself
A USB cable is included for data transfer
The software can also be used for other Lindab instruments, and a more detailed description is available separately

13. Operation and maintenance

There are user serviceable parts inside the LT 600 The device should never be opened by the user
Only specially trained personnel can open the device.
CAUTION - DANGER TO LIFE! max. 230V 50Hz
The device requires no maintenance apart from the occasional application of light grease on the round sealing ring on
the 50 mm connections (Silicone grease included)
To change the main fuse, unplug the device from the mains and pull the fuse holder out from the upper edge The ne-
wire fuses must only be replaced with another fuse of the same type
The measuring accuracy and operation of the device should be checked regularly (annual checks are suggested) at
the factory or by a suitably equipped testing centre
Figure 5: Nameplate and device no
The device should always have a filter pad fitted in the air intake during operation.
The air intake and outlet must be protected from dirt and moisture It is essential to prevent dust and water entering
the unit
Replace the lter pad on a regular basis A loss of power may indicate dirt at the intake
The device should only be run from stable electricity networks, not from generators or other supplies that are unable
to deliver continuous power
The LT 600 has been approved for use as a measuring instrument It should not ordinarily be used as a way of locating
leakages in duct systems, a process that can sometimes take hours However, if it is necessary to maintain pressure
for an extended period, you can avoid overloading the fan unnecessarily by not using an adapter
If the intake is used to try to locate a leakage, no smoke cartridges or mist of any kind may be used.
There is a risk of damaging the device.
17

Publicité

Table des Matières
loading

Table des Matières